Request Information Download PDF
GTM91110P240xx GTM91110P240xx
Information
Model NumberGTM91110P240xx
DescriptionGTM91110P240xx, Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC, , Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 240 Watts, 12-55V in 0.1V increments, Approvals:
Agency Documentshttp://www.globtek.info/certs/GTM91110P/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000Bz790EAB
CE EC-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
RoHS/RoHS2 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0Oa000000Bz790EAB
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC
CategoryMedical Power Supply
Input Voltage100-240V~, 50-60 Hz
I/P Amps (A)4.0
Wattage (W)240
Vout Range (V)12-55
Efficiency
Ingress Protection
Size (mm)127*76*43
EnclosureUser-added image
Rating Table
Model Number VoltageAmps (A)Watts (W) RFQ
GTM91110P24012-FA-S 1220240 RFQ
GTM91110P24015-FA-S 1516240 RFQ
GTM91110P24018-FA-S 1813.333239.99 RFQ
GTM91110P24024-FA-S 2410240 RFQ
GTM91110P24036-6.0-FA-S 308240 RFQ
GTM91110P24036-FA-S 366.66239.76 RFQ
GTM91110P24048-FA-S 485240 RFQ
GTM91110P24055-FA-S 554.36239.8 RFQ
SpecificationsA) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated 100-240 Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50/60 Hz
03. Inrush current: Cold start @ 25°C, 40A @ 115Vac, 80A @ 230Vac
04. Efficiency: 85% to 90% typical at full load, 120-240Vac input
05. Power factor: 0.98 typical at full load, 230Vac input
06. Turn on delay: 5 Seconds maximum, (3 seconds max option)
07. Load regulation: ± 1% Measured at the connector side
08. Line regulation: ± 0.2% Measured at the connector side
09. Output Ripple protection (Peak to Peak): +/-1% maximum of nominal output voltage measured at 20 MHz bandwidth with 0.1µf ceramic capacitor in parallel with a low impedance 47 µf electrolytic capacitor located 200mm from the end  of the output connector
10. Turn on/off overshoot: 5% Maximum, 1mS recovery time
11. Hold up time: 8 mS @ 300W load, 10mS@200W load, 16mS@150W load at 115Vac input
12. Transient response: 2% Maximum deviation, returns to initial condition in 3 ms max for 25% load step,   from 10% to 100% of rated load
13. Remote sense (optional): compensates for up to 0.5V voltage drop in output wiring.
Power supply will operate correctly, with or without remote sense pins hooked up.
14. Auxillary Output: For forced air fan 12V @ 250mA, if convection cooled 12V @ 150mA. Other output voltages and current ratings from 5V to 24V available, contact factory for availability.
15. Indicator: Green LED for output presence

B) PROTECTION
01. Input Protection: Input line fusing and 300Vac MOV
02. Short circuit: Protected. Unit will autorecover upon removal of short
03. Overload protection:  Autorecover upon removal of overload
    150W, 175% - 210% of nominal power rating
    200W, 130% - 160% of nominal power rating
    240W, 110% - 130% of nominal power rating
    300W, 110% - 130% of nominal power rating
04. Over voltage: Protected at 110- 140% of nominal output voltage, latched type. AC cycling required to restore output. Autorecovery type OVP is an option, contact factory for availability.
05. Fan output protection: 3 Pin Linear Regulator, short circuit proof

C) SAFETY
01. Dielectric Withstand Voltage: 4000Vac or 5656Vdc from input to output,
     On Class I models, 2500Vac or 3535Vdc from input to earth
02. Earth Leakage Current:
     Class I Input Models: Does not exceed 500uA
     Class II Input Models: Does not apply
03. Touch Current: Does not exceed 100uA, output terminations to earth
04. Output terminations rated as SELV, and Means of Operator Protection

D) OTHER:
01. MTBF: 200,000 Hours @ 25°C ambient temperature
02. Operating Temperature Range:
        -10°C to 50°C @ 100% output power rating
        -10°C to 70°C @ 50% output power derating
03. Cooling:
          Open Frame 150W Series: Convection
          Open Frame 240W and 300W Series: 20 CFM Forced Air Cooling
          Perforated Plastic Housing 200W Series: 20 CFM Forced Air Cooling,
                   with convection cooling 50% derated (100W) output capability.
04. Relative Humidity: operating 0% to 90% Non-condensing
05. Storage Temperature: -30°C to 85°C
06. Altitude: up to 2000 meter
07. Random vibration/transient drop: MIL-STD 810E

E) MECHANICAL  FORMAT:
150W, 240W and 300W Open Frame version
      Size: 127*76.2*43 mm
      Weight: 470 grams
200W Perforated Plastic Enclosure version
      Size: 140*87.6*51.9 mm
      Weight: 620 grams

 
Derating CurveUser-added image
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)135 x 85 x 60mm
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)30.0
Label Height (mm)20.0
Input Configuration
Description
Popular Products
SKUVolts OutCurrent Out (Amps)Connector
1732111216.7Stripped 7 +/- 1mm
1581681416P/N: 09-50-3041, KK® 3.96mm Crimp Terminal Housing, Friction Ramp, 4 Circuits or equivalent
1739711417.14MOLEX 6-PIN FRICTION LOCK HEADER P/N 26-60-4060 MATES WITH PN: 09-50-3061
1840431416P/N: 09-50-3041, KK® 3.96mm Crimp Terminal Housing, Friction Ramp, 4 Circuits or equivalent
1511872410P/N: 09-50-3041, KK® 3.96mm Crimp Terminal Housing, Friction Ramp, 4 Circuits or equivalent
225335554.36P/N: 09-50-3041, KK® 3.96mm Crimp Terminal Housing, Friction Ramp, 4 Circuits or equivalent

问询